MedtronicCVEs & Vulnerabilities

29 CVEs affecting Medtronic products, tracked from the National Vulnerability Database, with CVSS/EPSS scores and exploitation status.

Most Affected Products

valleylab ft10 energy platform firmware 6valleylab ft10 energy platform 4carelink network 4paradigm wireless insulin pump 42090 carelink programmer 3mycarelink smart model 25000 32090 carelink programmer firmware 3mycarelink smart model 25000 firmware 3
CVE-2025-12997LOW

Insecure Direct Object Reference vulnerability in Medtronic CareLink Network which allows an authenticated attacker with access to specific device and user information to submit web requests to an API endpoint that would expose sensitive user information. This issue affects CareLink Network: before December 4, 2025.

4 Dec 2025
3.1
CVSS
CVE-2025-12996MEDIUM

Medtronic CareLink Network allows a local attacker with access to log files on an internal API server to view plaintext passwords from errors logged under certain circumstances. This issue affects CareLink Network: before December 4, 2025.

4 Dec 2025
4.1
CVSS
CVE-2025-12995CRITICAL

Medtronic CareLink Network allows an unauthenticated remote attacker to perform a brute force attack on an API endpoint that could be used to determine a valid password under certain circumstances. This issue affects CareLink Network: before December 4, 2025.

4 Dec 2025
9.8
CVSS
CVE-2025-12994MEDIUM

Medtronic CareLink Network allows an unauthenticated remote attacker to initiate a request for security questions to an API endpoint that could be used to determine a valid user account. This issue affects CareLink Network: before December 4, 2025.

4 Dec 2025
5.3
CVSS
CVE-2023-31222HIGH

Deserialization of untrusted data in Microsoft Messaging Queuing Service in Medtronic's Paceart Optima versions 1.11 and earlier on Windows allows an unauthorized user to impact a healthcare delivery organization’s Paceart Optima system cardiac device causing data to be deleted, stolen, or modified, or the Paceart Optima system being used for further network penetration via network connectivity.

29 Jun 2023
8.8
CVSS
CVE-2023-25931MEDIUM

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

1 Mar 2023
6.8
CVSS
CVE-2022-32537MEDIUM

A vulnerability exists which could allow an unauthorized user to learn aspects of the communication protocol used to pair system components while the pump is being paired with other system components. Exploitation requires nearby wireless signal proximity with the patient and the device; advanced technical knowledge is required for exploitation. Please refer to the Medtronic Product Security Bulletin for guidance

12 Dec 2022
4.8
CVSS
CVE-2020-27252HIGH

Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.

14 Dec 2020
8.1
CVSS
CVE-2020-25187CRITICAL

Medtronic MyCareLink Smart 25000 is  vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device

14 Dec 2020
9.8
CVSS
CVE-2020-25183HIGH

Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

14 Dec 2020
8.8
CVSS
CVE-2019-13543HIGH

Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use multiple sets of hard-coded credentials. If discovered, they can be used to read files on the device.

8 Nov 2019
7.5
CVSS
CVE-2019-13539HIGH

Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use the descrypt algorithm for OS password hashing. While interactive, network-based logons are disabled, and attackers can use the other vulnerabilities within this report to obtain local shell access and access these hashes.

8 Nov 2019
7.8
CVSS
CVE-2019-13535MEDIUM

In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data.

8 Nov 2019
4.6
CVSS
CVE-2019-13531MEDIUM

In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.

8 Nov 2019
4.6
CVSS
CVE-2019-10964HIGH

Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

29 Jun 2019
7.1
CVSS
CVE-2019-6540MEDIUM

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.

26 Mar 2019
6.5
CVSS
CVE-2019-6538MEDIUM

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

26 Mar 2019
6.5
CVSS
CVE-2018-18984MEDIUM

Medtronic CareLink and Encore Programmers do not encrypt or do not sufficiently encrypt sensitive PII and PHI information while at rest .

14 Dec 2018
4.6
CVSS
CVE-2018-10634MEDIUM

Communications between Medtronic MiniMed MMT pumps and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.

14 Aug 2018
4.8
CVSS
CVE-2018-10626MEDIUM

Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.

10 Aug 2018
4.4
CVSS
CVE-2018-10622MEDIUM

Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication.

10 Aug 2018
5.2
CVSS
CVE-2018-10631MEDIUM

The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.

13 Jul 2018
6.3
CVSS
CVE-2018-8870MEDIUM

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.

3 Jul 2018
6.4
CVSS
CVE-2018-8868MEDIUM

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

3 Jul 2018
6.2
CVSS
CVE-2018-10596HIGH

Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.

3 Jul 2018
7.1
CVSS
CVE-2018-8849MEDIUM

Medtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programme and 8870 N'Vision removable Application Card do not encrypt PII and PHI while at rest.

18 May 2018
4.6
CVSS
CVE-2018-5448MEDIUM

Medtronic 2090 CareLink Programmer’s software deployment network contains a directory traversal vulnerability that could allow an attacker to read files on the system.

4 May 2018
4.8
CVSS
CVE-2018-5446MEDIUM

Medtronic 2090 CareLink Programmer uses a per-product username and password that is stored in a recoverable format.

4 May 2018
4.9
CVSS
CVE-2011-3386MEDIUM

Unspecified vulnerability in Medtronic Paradigm wireless insulin pump 512, 522, 712, and 722 allows remote attackers to modify the delivery of an insulin bolus dose and cause a denial of service (adverse human health effects) via unspecified vectors involving wireless communications and knowledge of the device's serial number, as demonstrated by Jerome Radcliffe at the Black Hat USA conference in August 2011. NOTE: the vendor has disputed the severity of this issue, saying "we believe the risk of deliberate, malicious, or unauthorized manipulation of medical devices is extremely low... we strongly believe it would be extremely difficult for a third-party to wirelessly tamper with your insulin pump... you would be able to detect tones on the insulin pump that weren't intentionally programmed and could intervene accordingly."

3 Sep 2011
4.0
CVSS