CVE-2018-10596
CWE-923Published: July 3, 2018· Updated: Jun 17, 2026
Official Description
Medtronic 2090 CareLink Programmer
uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
Technical Analysis
CVE-2018-10596 can be exploited remotely over the network without requiring physical or adjacent access, significantly expanding the attack surface for threat actors.
Exploitation requires low privileges, which limits the exposure to scenarios where an attacker has already gained initial access.
A successful exploit results in complete confidentiality breach (data exposure), full integrity compromise (data manipulation), availability disruption (denial of service), with a CVSS base score of 7.1.
CVSS v3.1 Vector Breakdown
Affected Vendors & Products
Exploit & PoC Resources
Official Patches & Advisories
All References (3)
Quick Facts
Related CVEs (CWE-923)
Recommended Actions
- →Apply vendor patches immediately
- →Monitor CVE-2018-10596 in threat intel feeds
- →Review IDS/IPS signatures for exploitation attempts