HOMEVULNERABILITIESCVE-2024-48966
CRITICAL

CVE-2024-48966

CWE-306Published: November 14, 2024· Updated: Jun 17, 2026

10.0
CVSS v3.1

Official Description

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

NVD Source

Technical Analysis

CVE-2024-48966 can be exploited remotely over the network without requiring physical or adjacent access, significantly expanding the attack surface for threat actors.

The vulnerability requires no privileges and no user interaction, making it a prime target for automated exploitation campaigns and worm-like propagation.

A successful exploit results in complete confidentiality breach (data exposure), full integrity compromise (data manipulation), availability disruption (denial of service), with a CVSS base score of 10.0.

The vulnerability has a "Changed" scope, meaning successful exploitation can impact components beyond the vulnerable component itself — such as the host operating system or adjacent services.

CVSS v3.1 Vector Breakdown

Exploitability
Attack VectorNetwork
Attack ComplexityLow
Privileges Req.None
User InteractionNone
ScopeChanged
Impact
ConfidentialityHigh
IntegrityHigh
AvailabilityHigh
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

Exploit & PoC Resources

NO KNOWN EXPLOITNo public exploit confirmed at this time
External links open in a new tab. Always verify in a controlled environment before use.

All References (1)

Quick Facts

CVE IDCVE-2024-48966
CVSS Score10.0 / 10
SeverityCRITICAL
WeaknessCWE-306
CISA KEVNo
PublishedNov 14, 2024

Related CVEs (CWE-306)

Recommended Actions

  • Apply vendor patches immediately
  • Monitor CVE-2024-48966 in threat intel feeds
  • Review IDS/IPS signatures for exploitation attempts
Data sourced from NVD (NIST), CISA KEV, and EPSS (FIRST). Analysis generated by CTIWATCH.COM. CVE data is provided under the NVD usage policy.