CVE-2019-25719
CWE-924Published: June 2, 2026· Updated: Jun 3, 2026
Official Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Technical Analysis
CVE-2019-25719 can be exploited remotely over the network without requiring physical or adjacent access, significantly expanding the attack surface for threat actors.
The vulnerability requires no privileges and no user interaction, making it a prime target for automated exploitation campaigns and worm-like propagation.
A successful exploit results in availability disruption (denial of service), with a CVSS base score of 8.6.
CVSS v3.1 Vector Breakdown