Official Description
Philips ISCV application prior to version 2.3.0 has an insufficient session expiration vulnerability where an attacker could reuse the session of a previously logged in user. This vulnerability exists when using ISCV together with an Electronic Medical Record (EMR) system, where ISCV is in KIOSK mode for multiple users and using Windows authentication. This may allow an attacker to gain unauthorized access to patient health information and potentially modify this information.
Technical Analysis
CVE-2018-5438 requires local access, meaning attackers must already have a foothold on the target system.
Exploitation requires low privileges, which limits the exposure to scenarios where an attacker has already gained initial access.
A successful exploit results in complete confidentiality breach (data exposure), full integrity compromise (data manipulation), with a CVSS base score of 6.3.