CVE-2015-3954
CWE-285Published: March 25, 2019· Updated: Jun 17, 2026
Official Description
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Technical Analysis
CVE-2015-3954 can be exploited remotely over the network without requiring physical or adjacent access, significantly expanding the attack surface for threat actors.
The vulnerability requires no privileges and no user interaction, making it a prime target for automated exploitation campaigns and worm-like propagation.
A successful exploit results in complete confidentiality breach (data exposure), full integrity compromise (data manipulation), availability disruption (denial of service), with a CVSS base score of 9.8.
CVSS v3.1 Vector Breakdown
Affected Vendors & Products
Exploit & PoC Resources
Official Patches & Advisories
All References (2)
Quick Facts
Related CVEs (CWE-285)
Recommended Actions
- →Apply vendor patches immediately
- →Monitor CVE-2015-3954 in threat intel feeds
- →Review IDS/IPS signatures for exploitation attempts